Magtrol Model 5300 Three Phase Power Analyzer
Chapter 4 - Operation
DISPLAY SELECTIONS
Figure 8. Display Selections
Ø1 Ø2 Ø3 K
W
Σ
DISPLAY Ø1, Ø2, Ø3
Individually selects Phase 1, Phase 2 or Phase 3
WATTS for display.
If all three push-buttons are simultaneously pressed, the
four LEDs illuminate, indicating that the AMPS display
reads the SUM of the three phase currents, the VOLTS
reads the AVERAGE of the three phase voltages and
the WATTS reads SUM of the three phase powers, as
follows:
AMPERES I I I=++
φφ φ
123
(SUM)
VOLTS
VVV
=
++
φφ φ
123
3
(AVERAGE)
KILOWATTS
WWW
=
++
φφ φ
123
1000
(SUM)
DISPLAY
ΣΣ
ΣΣ
ΣK
W
The WATTS display indicated the total 3-phase power
in kilowatts.
KW
W W W
=
+ +
φ
φ
φ
1 2 3
1000
(SUM)
VOLTAGE RANGES
Figure 9. Voltage Range Selections
150 300 600 AUTO
150, 300 and 600 volts rms plus AUTO range. At power
turn on, the 150 volt range and AUTO (auto-ranging)
are active, by default. Up ranging occurs when the
measured voltage on any phase exceeds that range
maximum voltage by 5% (>105%). This feature is active
in either AUTO (LED on) or manual (LED off).
Automatic down ranging occurs only in AUTO when
the measured voltage falls to about 1/4% below that
range maximum value.
If MODE PH. V. is off, up ranging occurs at 105% of
the range times
3
.
The PA up-ranging feature protects the instrument from
overload and ensures accurate measurements.
Exceeding the PA maximum voltage of 600 volts causes
the instrument to display "HELP." This means that
you must reduce your input voltage.