Magtrol Model 5300 Three Phase Power Analyzer
Chapter 9 - Calibration
AMPERES CALIBRATION CHECK
DC:
1. On the 5300, press the 5 AMPERES range push-
button.
2. Set the current calibrator for zero amperes and
operate mode. The AMPERES display should read
zero + 0.0001% max of range full scale.
3. Set the current calibrator for + 5 Amperes DC. The
AMPERES display should read 5.000 Amps.
Record the current readings.
4. Set the current calibrator for - 5 Amperes. The
AMPERES display should read 5.000.
5. Repeat the above for the 25 and 100 Amp ranges.
6. Compare the above readings to the values given
in Section 2 - Specifications.
The (+) and (-) readings should be within ± 2 least
significant digit (LSD) of each other.
AC:
1. Set the current calibrator for 60 Hz. and repeat the
current checks outlined above.
NOTE: If a beat frequency effect is
experienced between the
calibrator output and the 60 Hz
power line frequency, try setting
the calibrator for 80 Hz instead of
60 Hz.
2. Set the current calibrator to standby mode.
WATTS CALIBRATION CHECK
This step calls for the simultaneous application of
voltage and current to the 5300 PA.
DC:
1. On the 5300, select a VOLTS and AMPERES
range.
2. Set the volts and current calibrators for a voltage
and current near the center of each selected range.
Switch the calibrators to OPERATE mode.
3. The WATTS readout should show a power that is
the product of the VOLTS and AMPS indications
( POWER = V × I ).
4. Compare the 5300 PA WATTS readout to the
product of VOLTS and AMPS from the calibrator's
output.
This difference is the WATTS error.
AC:
POWER = V × I × Cos θ
where = phase angle between volts and amps.
NOTE: Maximum true power is
delivered only when θ = zero
degrees (cos θ = 1). Therefore,
the volts calibrator variable phase
adjustment is set for zero phase
between the volts and amps at the
input to the 5300 PA. Refer to
the calibrator instruction manuals
for this adjustment procedure.
The phase shift control should be carefully set to provide
a peak 5300 PA WATTS indication, corresponding to
zero volts to amps phase angle. Repeat the procedure
as used for the DC power verification and record
readings.
Compare the recorded reading to the values given in
Section 2 - Specifications.
MINOR ADJUSTMENT (IF NEEDED)
NOTE: All trim pot adjustments are
located on the 78B146 and
78B147 circuit boards of the
5300. See Figure 17 for trimpot
adjustment locations.
Lethal voltages are present. Take
appropriate safety precautions.